5.2 PRESSURE ZERO SETTING
Before you connect the pressure fitting to the PRU-353 you should set the zero
point of the pressure channel (channel 3). This setting should be made before
each recording run and compensates for changes in atmospheric pressure. With
the pressure port open to the atmosphere activate the bottom right switch ( 0).
After the zero has been set you can connect the pressure fitting to the port and
then pierce the container to let the pressure reach the sensor. The pressure port
(channel 3) is on the back edge of the instrument to the left of the display.

5.3 THERMOMETER

When the instrument is ready to record, the temperature of the probes is shown
on the display, updated every second.
as a highly accurate thermometer.

5.4 STARTING THE RECORDING

When you are ready to put the instrument into the pasteuriser, start recording by
activating the top right switch ( ).
*
The illustration shows the English language version in °C and bar. The text changes according to the language and
units selected for the display.