ATTENTION!
It is possible that an error could be caused by very severe interference from other nearby
equipment or on the mains wiring. If this is suspected then the cause should be investigated
because the levels of interference will have to be very high to cause errors.
Template Exchange
You will also see this message if there is a communications error while transferring the template
directory between two RPC-80 units when exchanging templates.
Set up the two RPC-80 units in the correct order, see chapter 8.5.3.
The lead that connects the two units may have become disconnected. Never unplug the
lead until the template exchange is completed.
Check that you are using the correct type of null modern and cable.
It is impossible to recover from this type of error and you will have to re-start the exchange from
scratch.
C2
it is impossible to communicate with the monitor to set up the active channels etc. This applies
when you are in the "Configuration" mode of operation.
The lead that connects the monitor to the RPC-80 may have become disconnected.
The monitor may not be switched on.
The monitor may be an 100 series type, only the 200 series monitors can be set up this
way.
You will also see this message if there is a communications error during certain stages
of the Cal-check
ATTENTION
!
Page 80 of 112
Never unplug this lead until the "Printing Finished" message shows and
the charge rate has reduced to trickle charge.
It is possible that an error could be caused by very severe interference
from other nearby equipment or on the mains wiring. If this is suspected
then the cause should be investigated because the levels of interference
will have to be very high to cause errors.