Connecting The Heated Sample Line With Special Adapter (Option); Electrical Connections; Figure 4 Heated Sample Line Connection With Special Adapter - M&C PSS Series Instruction Manual

Gas conditioning unit
Hide thumbs Also See for PSS Series:
Table of Contents

Advertisement

12.1.1 CONNECTING THE HEATED SAMPLE LINE WITH SPECIAL ADAPTER (OPTION)

Sample gas IN
Bulkhead union
Back ferrule 6 mm Ø ID
PSS-5
PSS-5/3
Support sleeve
4 mm Ø OD
Figure 4
Heated sample line connection with special adapter
• Place the special adapter on the PTFE tube according to the drawing seen above;
• Place the support sleeve into the PTFE tube;
• Insert the Teflon tube as far as possible into the 'Sample gas IN' bulkhead fitting and hand-tighten
the adapter;
• Tighten the adapter 1 1/4 turns with a wrench (SW 14), while holding the lock nut of the Schott
screw connection with a wrench (SW 15);
• Insert the 10mm pipe of the heating cable into the adapter as far as possible and hand-tighten with
the union nut;
• Tighten the union nut 1 1/4 turns with the wrench (SW 19), holding the adapter in place with the
wrench;
The screw connection is now cut gas-tight and can be loosened as often as required.
12.2

ELECTRICAL CONNECTIONS

W A R N I N G !
N O T E !
16
Adapter
Sealing ring
Sealing ring
6 mm Ø ID
10 mm Ø ID
False supply voltage can demage the equipment. When
connecting the equipment, please ensure that the supply voltage
is identical with the information provided on the model type
plate!
For the erection of power installations with rated voltages up to
1000V, the requirements of VDE 0100 and relevant standards and
specifications must be observed!
The main circuit is equipped with a fuse corresponding to the
nominal current (over current protection); for electrical details see
technical data.
PSS-5 | 1.00.03
Union nut 10 mm Ø ID
Back ferrule
10 mm Ø ID
www.mc-techgroup.com
Heated sample line