(IV) Voltage Settle Time
•
Set the time in seconds between applying a voltage and measuring the current.
I. This has a maximum of 10 seconds.
(V)
Illumination
•
Set the illumination intensity (in mWcm
(VI) Hysteresis I-V
•
This option performs a reverse current-voltage measurement after the forward current-voltage
measurement has completed.
I. This reverses the set start and end voltages and uses the same voltage increment and settle
time as the forward measurement.
9.5 Solar Lifetime Measurement Settings
9.5.1 Measurement Settings
(I)
Start Voltage
•
Set the voltage in volts at which to start the current-voltage measurement.
I. This can be set between -10 V and +10 V.
(II)
End Voltage
•
Set the voltage in volts at which to end the current-voltage measurement.
I. This can be set between -10 V and +10 V.
18
) being used during the measurement.
-2
Figure 9.8. Measurement Settings for the solar lifetime measurements.