Beckman Coulter AU680 Training Manual page 39

Chemistry analyzer
Table of Contents

Advertisement

Select Home > Rack Requisition Sample > Calibration
Review the automatic reagent blank and calibration
Select sample type from the Type drop-down list to view reagent blanks and
calibrations ordered (requisitioned) by the analyzer for each sample type
Notes:
Automatic order (requisition) occurs during a reagent check
Tests with a blue box are ordered (requisitioned) for reagent blank
Tests with a yellow box are ordered (requisitioned) for calibration
Select Start Entry F1
Select the RB or CAL column to the right of the test name to order
(requisition) reagent blank and calibration
Optional: Select Individual Requisition F3 to order (requisition)
sequenced bottles of the same test
Select sample type from the Type drop-down list to order (requisition)
reagent blank and calibration for other sample types
Select Entry F1 to save the order (requisition)
to view the reagent blank, calibrator(s), and racks required
Load the reagent blanks and calibrators according to the list in the correct
racks and place the racks on the rack supply component (blue rack first)
Select Start from the main button area
Review errors on the Error List in the Start dialog and perform any
corrective actions, if necessary
Select Start from the Start dialog
AU680 In-Lab Training Manual
Version 2.0 (September 2016)
Analyzer Calibration
order (requisition):
Automatic
order (requisition)
satisfactory?
Manually order (requisition) calibration:
Select Display Cup Set F5
Note: You may need to scroll down, or select Print,
to view all required calibrator racks
Yes
No
Beckman Coulter Professional Development
Page 39 of 152