11�4 Damping Function; 11�5 Peak Hold Function - IFM Electronic EFECTOR 600 TW2000 Operating Instructions Manual

Infrared temperature sensor
Table of Contents

Advertisement

If the diagnostics function is activated using the control keys on the instrument, it
will remain in this mode for 10 sec�
If the diagnostics function is not used, connect the diagnostics input (Pin
5) to minus� As an alternative, use a 4-pole cable box where Pin 5 is not
assigned�
11.4 Damping function
When the target object's temperature is erratic, the damping function smoothens
these temperature fluctuations in order to stabilize the measuring signal. The
greater the time constant [DAP], the lower the effect of these fluctuations on the
yielded temperature reading�
OUT2
[mA]
1: Output signal without smoothing function
2: Output signal with smoothing function
11.5 Peak hold function
It might often be desirable to determine the peak temperature during a defined
time period, for example when the objects to be measured move past the infrared
temperature sensor, resulting in temperature readings which would appear to be
cyclical� In this mode, the displayed temperature reading will not drop between tar-
geted objects� The peak temperature reading will be held for a preset time period�
The hold time [PhLd] can be set at 100, 300 and 500 msec� The maximum
temperature sampled during the defined hold time will be saved� It makes sense to
choose a hold time which is approximately 1�5 times as long as the cycle of the mov-
ing targets� This avoids temperature drops� Any changes are recognised at once�
1
2
UK
t
19

Advertisement

Table of Contents
loading

Table of Contents