Atmos Cam 21 Operating Instructions Manual

Hide thumbs Also See for Cam 21:

Advertisement

English
GA1GB.150201.0
2018-11 Index: 07
ATMOS
Endoscopy and
Stroboscopy
System
®

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Cam 21 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Atmos Cam 21

  • Page 1 English ATMOS ® Endoscopy and Stroboscopy System GA1GB.150201.0 2018-11 Index: 07...
  • Page 2: Table Of Contents

    Function ..............4 Explanation of symbols .........5 For your safety ............6 Setting up and starting up .........7 Construction ............7 Starting up ATMOS Cam 21 / 31 ......9 ® Digital monitor outlets ...........9 Starting up ATMOS Strobo 21 LED ....11 ® Cleaning and care ..........13 General information on cleaning and disinfection........13...
  • Page 3: Introduction

    ® besides regular cleaning. Repair work and period tests may be carried out only by expert personnel authorised by ATMOS. By applying only original spare parts you will have the guarantee that operational safety, readiness for work and the value of your ATMOS Endoscopy and Stroboscopy System will be preserved.
  • Page 4: Intended Use

    Application restrictions: No Sterility: unsterile Active medical product Please observe the separate operating instructions for ATMOS Cam 21 / 31 and ATMOS Strobo 21 LED. ® ®...
  • Page 5: Explanation Of Symbols

    Colour 3200 K Typ. angels: 0°, 30°, 45° temperature: Flexible Endoscope Extra-light source Usage: Nose, throat, larynx Stroboscopy for connection of an ATMOS ® Strobo 21 LED examinations Microscope LED light source Usage: Examination of ears Typical: 2 W (e.g. ATMOS ®...
  • Page 6: For Your Safety

    • Please do not perform any modifications of the ATMOS ® consumables. Endoscopy and Stroboscopy System without approval • ATMOS is not liable for personal injury and damage to by the manufacturer. Please observe general rules for property if accident avoidance as well as for environment protection.
  • Page 7: Setting Up And Starting Up

    Strobo 21 LED from below on the left ® side. The ATMOS Cam 21 / 31 you tighten on the right side. ® Screw the support for the LED light handle to the metal sheet on the left and the support for the camera head to the right.
  • Page 8 3.0 Setting up and starting up The power supply at the monitor is plugged in and fixed with the Velcro fastener below the metal sheet where camera and stroboscope are positioned. Connecting the SVHS cable Connecting the DVI cable...
  • Page 9: Starting Up Atmos ® Cam 21 / 31

    3.3 Digital monitor outlets Warning notice / safety note The cameras ATMOS Cam 31 DV and ATMOS Cam 21 DV have a digital monitor outlet with SXGA signal. The ® ® resolution is 1280 x 1024 pixels. Please make sure that only monitors are connected which support the stated resolution.
  • Page 10 Please pay attention, just to switch on the devices after they have been completely wired. If required, analogue peripheral video equipment (monitor, video printer, etc) must be adjusted to NTSC video norm. For further details please look at the corresponding operating instructions. Only valid for ATMOS ® Optional:...
  • Page 11: Starting Up Atmos ® Strobo 21 Led

    ATMOS ATMOS Cam 31 ® ® ® When applying the ATMOS Cam 21 / 31 for video ® 21 LED 21 / 31 DV Data stroboscopy please pay attention to connecting the dongle connector (Ref 507.4781.0) with the interlink connection of ATMOS Cam 21 / 31 ®...
  • Page 12 The 6pin interlink cable you plug in at the "interlink" jack. ATMOS Strobo 21 LED ATMOS Cam 21/31 ® ® Connect the audio cable to the audio channel of the stroboscope with the black connector. Please do not connect the red cable.
  • Page 13: Cleaning And Care

    4.1.1 Cleaning the unit surface • The surfaces of the ATMOS Endoscopy and ® Stroboscopy System are resistant against all surface disinfectants listed in on page 14.
  • Page 14: Recommended Instrument Disinfectants

    4.0 Cleaning and care 4.2 Recommended instrument disinfectants Disinfectant Ingredients (in 100 g) Manufacturer Gigasept FF succindialdehyde 11.0 g Schülke & Mayr, Norderstedt (Application concentrate) dimethoxytetrahydrofurane 3.0 g corrosion protection components non-ionic surfactants and perfumes Sekusept PLUS glucoprotamin 25.0 g Henkel, Düsseldorf (Application concentrate) non-ionic surfactants Solvents, complexing agents...
  • Page 15: Maintenance And Service

    Mounting the replacement castors is performed in reverse order. • Affix the envelope to the outside of the package. • Send the product to ATMOS or to your dealer. Exchanging drawers: Drawers can only be exchanged as complete components. Please observe the provided mounting instructions.
  • Page 16: Options, Accessories And Consumables

    Capture Suite HD license dongle 700.0047.0 ® ATMOS Cam 21 507.4100.0 ® ATMOS Cam 21 DV 507.4120.0 ® ATMOS Cam 21 DV Data 507.5100.0 ® ATMOS Cam 31 with image memory 507.5110.0 ® ATMOS Cam 31 DV with image memory 507.5130.0 ® ATMOS Cam 31 Data with image memory 507.5120.0...
  • Page 17  High resolution naso-pharyngoscope Ø 3,8 mm, working length 300 mm, 950.0222.0 0°, angle of field of view 80°, depth of focus: 5 mm – infinite, angle: 125° / 125° Scope of delivery includes aluminium transport case and leak tester Flexible video-nasopharyngoscope ATMOS Scope 950.0300.0 ® ATMOS Scope Basic 950.0330.0...
  • Page 18: Technical Data

    IX, EC Directive 93/42/EEC CE marking GMDN code 11585 UMDNS code 11-585 ID No. (REF) 950.0223.0 Endoscopy Trolley 950.0226.0 Stroboscopy Trolley Please observe the separate operating instructions for ATMOS Cam 21 / 31 and ATMOS Strobo 21 LED. ® ®...
  • Page 19: Checking / Disposal

    In order to guarantee a proper disposal of your old device, please either pass on your old device to your specialised dealer or send it to ATMOS MedizinTechnik for disposal. Before disposal respectively before transport all parts, which came into contact with the patient must be thoroughly cleaned, disinfected.
  • Page 20: Ludwig-Kegel-Straße

    ATMOS MedizinTechnik GmbH & Co. KG Ludwig-Kegel-Straße 16 79853 Lenzkirch / Germany Phone: +49 7653 689-0 atmos@atmosmed.de www.atmosmed.com...