Actual Load Calibration Procedure - Unipulse F159 Operation Manual

Plug-in weight module for the omron sysmac cji series plc
Table of Contents

Advertisement

4.CALIBRATION

4-2. Actual Load Calibration Procedure

Steps required to perform actual load calibration are as follows:
38
Toggle the DIP switch 1 to OFF position (the switch is
LOCK Release
located under the terminal block).
Write initial setting data to the DM area that corresponds
Input initial
setting data
to MACH No.
Register initial setting data to F159. To do this, either
F159 Power on,
power on the F159, or restart it by toggling restart flag in
or restart
the following sequence: OFF → ON → OFF.
Set Soft LOCK bit (13th bit of n CH) to "0".
Release Soft LOCK
Register initial zero value by setting the Zero calibration
Zero Calibration
bit (0th bit of n+1 CH) to "1".
Register span value by loading the cell with actual span
Span Calibration
load and setting the Zero calibration bit (0th bit of n+1
CH) to "1"
Perform the Zero calibration procedure again if required.
Zero Calibration
Inhibit calibration procedures to avoid unauthorized or
Soft LOCK
accidental alteration.
Toggle the LOCK switch to ON position to avoid
LOCK Switch ON
unauthorized or accidental modification of calibration.
Removal of the terminal block is required for this
operation: make sure that NOV RAM bit (6th bit of n+6
CH) is in "0" position and power off the F159 before
trying to detach the terminal block.