Measure The Sample - Malvern Mastersizer Series Getting Started

Table of Contents

Advertisement

'
N O T E
If you are not in the
Measure-background
dialogue you can select the
Measure-inspect dialogue
by selecting Inspect from
the Measure menu or
alternatively you can select
the Inspect button.
%
$
2
. Press the Start button and the instrument will start to measure the obscura-
tion. The Start button will change to "Stop" while the measurement takes
place, allowing you to stop the measurement if you need to.
. Using the pipette add the sample to the dispersion unit. Add a few drops at
a time and allow the sample to be thoroughly mixed within the dispersant.
Look at the obscuration bar. The objective is to add enough sample to turn the
bar green (giving a value of between 10 and 30%) on the readout Á.
Once you are satisfied that the sample concentration is in the correct range you
can continue to the final stage of the measurement.
Caution
Do not spill any dispersant or sample onto the surfaces of the cover. It has been
found that certain substances can cause permanent damage to the covers. All
spillage should be scrupulously cleaned up immediately.

Measure the sample

The final stage of the measurement procedure is to measure the sample.
To measure the sample:
. From the Measure-Inspect dialogue press the Next button. The dialogue
below will appear.
1
G E T T I N G S T A R T E D
C H A P T E R 4
P a g e 4 . 1 3