C H A P T E R 1 0

Blending results

P a g e 1 0 . 8
G e t t i n g
S t a r t e d
30
20
10
0
The example above shows the result of transforming a skewed volume
distribution À to number Á.
In some cases the sample size range may be so broad that it cannot be covered by a
single range lens but can be covered by the combination of two range lenses. The
results measured by the two range lenses can be blended to produce a result with
a broader size range.
The blending takes place within the overlapped region of the two size ranges. The
two results within this region are combined to give a smooth transaction of result
from one size range to the other.
In the following example the sample has a size ranging from 1 micron to about
3000 microns. When measured with the 300mm range lens of the Mastersizer S
the results above 900 microns are cut off; whereas with a 1000mm lens the results
below 4 microns are cut off. To get a result covering the entire sample size range,
the results measured with the two range lenses are blended.