Malvern Mastersizer Series Getting Started page 177

Table of Contents

Advertisement

A P P E N D I X G
P a g e G . 4
G e t t i n g
S t a r t e d
. Complies with BS EN55022 (1995), class B, radiated and conducted emis-
sions.
. Complies with BS EN50082-2 (1995), generic immunity standard (indus-
trial environments).
. Complies with BS EN61000-4-2 (1995), Electrostatic discharge, to severity
level 4, using 8kV contact discharge and 15kV air discharge, with perform-
ance criteria 1, i.e. normal performance within specified limits, i.e. generic
performance criteria A
. Complies with DDENV50140, (1994), Immunity to radiated electromag-
netic energy to severity level 3 (10V/m 1kHz 80% AM 80MHz - 1GHz and
200Hz 1000PM at 900Hz) with performance criteria 1 i.e. normal perform-
ance within performance criteria 1. i.e. generic performance criteria A.
. Complies with BS ENV61000-4-4, immunity to electrical fast tran-
sient/bursts, to severity level 4 using 2kV on signal lines and 4kV on power
supplies, with performance criteria 1, i.e. normal performance within speci-
fied limits , i.e. exceeds specified requirements to generic performance crite-
ria A.
. Complies with BS ENV50141 (1995), Immunity to conducted RF distur-
bances to severity level (10Vrms 1KHz 80% AM 150KHz - 80MHz) with
performance criteria 1 i.e. normal performance within specified limits, i.e.
generic performance criteria A.
M A N