Convenient Functions - BRASSELER USA EndoDrive PLUS Operation Manual

Motor handpiece system for root canal enlargement
Table of Contents

Advertisement

AUTO REVERSE OFF
The motor handpiece stops without reverse rotation. In this instance, the LCD panel shows "
rotation speed alternately.
NOTICE
• When the reverse rotation (R) is set, the Auto Reverse Function is not activated.
• When the batteries go down (when the mark of the remaining amount of the batteries indicates "
the actual load may not reach the preset torque limit value. In this case (in the case of operation by batteries),
this auto reverse function will not be activated. When high torque is required, use the AC adapter or use this
product in a condition where the batteries are not significantly consumed (when the mark of the remaining
amount of the batteries indicates "
• If a load is continuously applied to the motor handpiece, it may automatically stop to prevent overheating. (The
torque limit value indicates "
cools down. When the display returns to normal, the motor handpiece can be used.
(6) Completion of Medical Treatment
When the treatment is completed, return the motor handpiece to the handpiece stand and hold down the POWER
key for more than one second to turn off the power.
NOTICE
If 10 minutes pass without operation of the ON/OFF button or Foot Control (Optional), the power turns o
automatically. (Auto power o function) However, the power does not turn o automatically if the motor
handpiece is in motion, even if it is not used.

6. CONVENIENT FUNCTIONS

(1) Program
You can change any preset values and have them memorized to your desired settings (rotation speed, torque limit
value, gear ratio and auto reverse mode).
1) Press the PROGRAM key until it turns to the program number which you want to have memorized.
2) Adjust the rotation speed, torque limit value, gear ratio and auto reverse mode by each key according to your
needs.
3) Hold down the MEMORIZE key for more than one second. When the alarm sounds, memorization is completed.
NOTICE
• The program cannot be memorized while the motor handpiece is in motion.
• The rotational direction is not memorized in the program. The program cannot be memorized if the rotational
direction is set to the reverse rotation (R).
• If the PROGRAM key is kept pressed, it is fast-forwarded.
• The program is not memorized unless the MEMORIZE key is held down for more than one second. If the
program number is changed by the PROGRAM key without pressing the MEMORIZE key, the initially
memorized preset values remain. (Cancel function)
(2) Initialization of Program (Factory-set configuration)
The program can be returned to the original state configured at the time of factory shipment, if setup becomes
confusing.
1) Turn off the power, when the power is on.
2) Connect the AC adapter and confirm that the AC IN lamp lights up.
3) Hold down the POWER key for more than one second, while pressing the AUTO REVERSE key and MEMORIZE key
at the same time.
4) Release the keys when the LCD panel indicates "
5) Initialization is completed, when "
,
").
" on the LCD panel). In this case, leave the motor handpiece for a while until it
", and re-press the MEMORIZE key while this is displayed.
" is displayed.