CAUTION
Do not apply high-temperature sterilization methods such as autoclave sterilization to any part other than the contra
angle head.

9. CHANGING BATTERIES

The motor handpiece uses rechargeable batteries. They can be recharged 300-500 times, depending on the use conditions
of the motor handpiece. If the operating time or recharging time becomes shorter or the rotation power becomes weaker,
although the "memory effect" described in "
case, ask your dealer to replace the batteries or replace them with new ones yourself. (Refer to the "
When replacing them yourself, be sure to observe the following "
Brasseler USA shall not be held liable for any malfunction or failure resulting from your not following the
"
CAUTIONS ON CHANGING BATTERIES."
* we recommend our genuine Brasseler USA Batteries.
Order Code : U421070
CAUTIONS ON CHANGING BATTERIES
• Do not open any part other than the battery cover.
• Be sure to purchase and use only the recommended batteries. Otherwise, batteries may cause damage, fluid
leakage or explode.
• Do not work with wet hands. This may cause trouble due to short-circuiting of batteries and moisture infiltrating
this product.
Changing Batteries
1) Turn off the power. Remove the AC adapter.
2) Remove the batteries after sliding the battery cover on the
back of the control unit in the direction of the arrow. (Fig. 13)
CAUTION
Make sure to remove the AC adapter before
changing batteries. Otherwise, it may cause a
malfunction.
3) Take out the old batteries, and pull out the cord, holding it
at the connector. (Fig. 14)
CAUTION
Make sure to pull out the cord by holding the connector.
Otherwise, it may cause a break in the cord.
4) Insert the connector of the cord of new batteries into the
connector in accordance with the polarity indication label
inside the battery chamber, and place the batteries into the
chamber with care so as not to catch the cord.
CAUTION
In case of difficult to insert the connector onto the battery, the polarity may be incorrect. Do not insert it by force.
NOTICE" is not applicable, the life of the batteries is depleted. In such a
CAUTIONS ON CHANGING BATTERIES." Note that
15
Battery Cover
Battery Cover
Connector
Connector
Changing Batteries.")
Fig. 13
Fig. 14