Error Code - BRASSELER USA EndoDrive PLUS Operation Manual

Motor handpiece system for root canal enlargement
Table of Contents

Advertisement

5) Close the battery cover. Make sure to charge the batteries before use.
NOTICE
The used nickel metal hydride batteries are recyclable, but their disposal may sometimes not be permitted by local
laws. Return them to Brasseler USA.

10. ERROR CODE

If the motor handpiece stops due to an abnormality such as a malfunction, overload, break or wrong use, it automatically
checks the state of the control unit and detects the cause of the abnormality and displays an error code on the LCD panel. If
an error code is displayed, turn on the power again and check whether the same error code is displayed. If the same error
code is displayed, take action by referring to the instructions provided in the "Check/Remedy" column in the following table.
NOTICE
If batteries are replaced while the AC adapter is connected, an incorrect error code may be displayed.
Error code
During
rotation of
the motor
handpiece
At the time
of charging
At the time
of calibration
Other
Error
Self-Check error
Malfunction of circuit
The motor handpiece is
Ovecurrent
locked. (at the time of the
auto reverse mode)
Overvoltage
Malfunction of circuit
High load was continuously
applied to the motor
Overheating of motor
handpiece for a relatively long
time.
Error of the brake
Malfunction of circuit
circuit
The motor handpiece is
The axle locks up
locked. (at the time of startup)
The voltage of batteries is too
Low voltage of
low (The life of batteries or not
batteries
iserted.)
High voltage of
The voltage of batteries is too
batteries
high. (Malfunction of circuit)
Beyond the range of working
Beyond the range of
temperature or break in the
working temperature
thermistor in the battery
section.
The operating life of the motor
Beyond the upper limit
handpiece or contra angle
Below the lower limit
has expired.
Abnormal heat
The batteries generate
generation from
abnormal heat.
batteries
Cause
16
Check/Remedy
Contact Brasseler USA.
Remove load
Contact Brasseler USA.
Leave it as it is until it cools down.
Contact Brasseler USA.
Remove load
Put the batteries into the battery
chamber, or replace with new
batteries.
Contact Brasseler USA.
Use within the range of working
temperature, or replace the batteries.
Replace the motor handpiece or
contra angle head.
Replace the batteries. If the heat
generates from the new set of batteries,
malfunction of the circuit may be
suspected. Contact Brasseler USA.