Trouble Shooting - Elma Elmasonic S Series Service Manual

Hide thumbs Also See for Elmasonic S Series:
Table of Contents

Advertisement

5
Improper use
Proceeding faulty
unit
Trouble-shootings
LED display
malfunction
Ultrasound
malfunction
Temperature
malfunction
© Elma Schmidbauer GmbH

Trouble shooting

Find out if the malfunction is caused by a fault on the unit or by
improper use. Examples for malfunctions due to improper use:
The cleaning result is not satisfactory:
The cleaning medium, the cleaning temperature or the
duration of the cleaning cycle are not suitable for the
cleaning task.
The cleaning result can also be impaired by a wrong or
insufficient filling level and/or very dirty cleaning liquid.
The unit does not heat up correctly:
For an efficient heating up of the cleaning liquid switch on
the ultrasound (see section 4.5. Start the ultrasound –
temperature-controlled*;with revolution of the cleaning bath)
and use a cover.
Check housing and bottom plate of the unit for any visible
damage and check the tank for leaks before you connect the
unit to an interrupting transformer.
Then connect the unit to an adjustable interrupting transformer
for a functional check. You can easily and safely check the unit
for a short circuit by slowly adjusting the mains voltage upwards
and observing the current consumption. At the same time you
can find out which functions of the unit have broken down.
Follow to the trouble-shootings in the Elmasonic S service-
pack.
Various malfunctions are indicated by a flashing LED display:
S 10 (H): LED for ultrasonic activity flashes fast:
S 15 – S 900 H: LED display cleaning period flashes fast
(„running light") / LED display in the ultrasound key flashes fast:
There is a fault on the PCB control, on the connection between
ultrasound >> transducer system or on the transducer system.
S 10 (H): LED display for heating operation flashes fast.
S 15 – S 900 H: LED display heating temperature flashes fast
(„running light") / LED display in the ultrasound key flashes fast:
There is a fault in the temperature measuring, e.g. interrupted
connection to temperature gauge or a faulty gauge.
SM_Elmasonic_S_02.2018_EN
Trouble shooting
19

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents