Trouble shooting

11
Fault
Casing damaged
Mains cable damaged
No operating functions; all
LEDs dark
No ultrasonic function;
LED ultrasound dark
No ultrasonic operation;
LEDs of LED cleaning
period blink alternately
("running light") = fault
indication ultrasound
Unsatisfactory cleaning
results
34
Trouble shooting
Possible cause
damage by third party,
transport damage
damage by third party,
transport damage
mains cable not plugged in
socket dead
mains cable
damaged/interrupted
fault of electronics
turning knob for ultrasonic
operation in „0" position
unit is switched off
key ►
(ultrasound) not
pressed
fault of electronics
fault of electronics
no or unsuitable cleaning
medium used
cleaning temperature not
sufficient
cleaning period too short
BA_Elmasonic S_EN_Vers. 05.2019
Remedy
return unit to supplier or
manufacturer
obtain original spare mains
cable from manufacturer or
supplier
plug in mains cable
check socket/fuse
replace mains cable
return unit to supplier or
manufacturer
switch on the turning knob
for ultrasonic operation
switch on the unit at key
on/off
press key ►
return unit to supplier or
manufacturer
switch unit off and on
if fault is indicated again:
return unit to supplier or
manufacturer
use suitable cleaning
medium
heat up cleaning liquid
repeat cleaning interval