... 8 Electrical connections
Electrical data for inputs and outputs
Input - resistance thermometer / resistances
Resistance thermometer
•
Pt100 in accordance with IEC 60751, JIS C1604, MIL-T-24388
•
Ni in accordance with DIN 43760
•
Cu in accordance with recommendation OIML R 84
Resistance measurement
•
0 to 500 Ω
•
0 to 5000 Ω
Sensor connection type
Two-, three-, four-wire circuit
Connection lead
•
Maximum sensor line resistance
per line 50 Ω in accordance with NE 89
•
Three-wire circuit:
Symmetrical sensor line resistances
•
Two-wire circuit:
Compensation up to 100 Ω total lead resistance
Measurement current
< 300 μA
Sensor short circuit
< 5 Ω (for resistance thermometer)
Sensor wire break
•
Measuring range: 0 to 500 Ω > 0.6 to 10 kΩ
•
Measuring range: 0 to 5 Ω > 5.3 to 10 kΩ
Corrosion detection in accordance with NE 89
•
Three-wire resistance measurement > 50 Ω
•
Four-wire resistance measurement > 50 Ω
Sensor error signaling
•
Resistance thermometer:
Sensor short circuit and sensor wire break
•
Linear resistance measurement:
Sensor wire break
Input - thermocouples / voltages
Types
B, E, J, K, N, R, S, T in accordance with IEC 60584
•
•
U, L in accordance with DIN 43710
•
C, D in accordance with ASTM E-988
Voltages
•
−125 to 125 mV
•
−125 to 1100 mV
Connection lead
•
Maximum sensor line resistance:
per line 1.5 kΩ, total 3 kΩ
Sensor wire break monitoring in accordance with NE 89
•
Pulsed with 1 µA outside measurement interval
•
Thermocouple measurement 5.3 to 10 kΩ
•
Voltage measurement 5.3 to 10 kΩ
Input resistance
> 10 MΩ
Internal reference junction Pt1000, IEC 60751 Cl. B
(no additional jumpers necessary)
Sensor error signaling
•
Thermocouple:
Sensor wire break
•
Linear voltage measurement:
Sensor wire break