Testing A Sample - Manual Clemen Unit; Adjusting The Tool - Elcometer 3000 Series Operating Instructions Manual

Clemen unit
Hide thumbs Also See for 3000 Series:
Table of Contents

Advertisement

3 TESTING A SAMPLE - MANUAL CLEMEN UNIT

Before the sample can be tested, adjust the height of the tool:

3.1 Adjusting the tool

1. Place the sample onto the sample platform and fix in position by rotating the
clamp handle.
2. Lift and hold the load beam in its raised position and slide the sample
platform until the centre of the sample is underneath the tool (Figure 3).
3. Insert the metal shim between the nylon wheel (1) and the load beam and
then lower the load beam.
2
4. Adjust the height of the tool (2) until it is touching the surface of the sample
and then tighten the knurled wheel (3) to lock the tool in position.
5. Raise the load beam, remove the metal shim and slide the sample platform
back to its starting position.
6. Slide the weight along the load beam to the required load and lock in place
by tightening the knurled wheel. The load can be set from 0 g to 2000 g
(0 lb to 4.4 lb).
6
R
3
Figure 3. Adjusting the scratching tool
1

Advertisement

Table of Contents
loading

This manual is also suitable for:

3000/13000/3

Table of Contents