Testing A Sample - Motorised Clemen Unit; Adjusting The Tool - Elcometer 3000 Series Operating Instructions Manual

Clemen unit
Hide thumbs Also See for 3000 Series:
Table of Contents

Advertisement

4 TESTING A SAMPLE - MOTORISED CLEMEN UNIT

Before the sample can be tested, adjust the height of the tool:

4.1 Adjusting the tool

1. Connect the unit to the mains supply and press the Power On/Off button.
2. Place the sample onto the sample platform and fix in position by rotating the
clamp handle.
3. Lift and hold the load beam in its raised position.
4. Press the Start button (Figure 2). Wait until the tool is halfway through its
travel and then press the Power On/Off button to stop the tool in this
position.
5. Insert the metal shim between the nylon wheel (1) and the load beam and
then lower the load beam.
2
6. Adjust the height of the tool (2) until it is touching the surface of the sample
and then tighten the knurled wheel (3) to lock the tool in position.
7. Raise the load beam, remove the metal shim and while supporting the
beam, press the Power On/Off button and then the Start button to return the
tool to the starting position.
Supporting the load beam while the tool is moving to its starting position
prevents damage to the sample surface, the tool or the sample platform.
8
R
3
Figure 5. Adjusting the scratching tool
1

Advertisement

Table of Contents
loading

This manual is also suitable for:

3000/13000/3

Table of Contents