5 Prepare the system for a run
5.4 Connect a column
Step
5
6
102
Action
In the Process Picture:
Click on the Column.
•
Select Column down flow.
•
Result: The Column valve switches to position 1.
In the Process Picture:
Click on the Pumps.
•
Enter a low System flow (e.g., 0.2 ml/min).
•
Click Set flow rate.
•
Result: A system flow of 0.2 ml/min starts.
ÄKTA pure Operating Instructions 29-0229-97 AE