April 2019
3

Installation

Note
Make sure to follow the instructions carefully for successful installation.
3.1
Verify reference reflector in kit matches application
Different reference reflectors are used depending on the installation
environment.
Procedure
Select reference reflector according to
Table 3-1: Application Categories
Device mounting type
3-in. to 8-in. pipe/chamber (inner
diameter)
10-in. pipe/chamber or bigger (inner
diameter) or Open tank (tank without
pipe)
Figure 3-1: Mounting Types
A
C
A. Pipe/chamber
B. Open tank
C. Inner diameter
Manual Supplement
Table
3-1.
Reference reflector type
Short 1 in. (25 mm)
Long 2.8 in. (70 mm)
B
Manual Supplement
Model code
HL1, HL2
HL3
7