Note:
Depending on the mA Operation Mode that is selected, if gas leak is present, the detector will indicate a gas leak SPL reading.
Note:
If the output reverts to 3 mA or < 1 mA, a Warning or Fault is present. Follow the instructions below to investigate and clear a Warning or
Fault (see Clear Faults or Warnings).
If local regulations or site practices require that a gas leak test is done to check that the detector is functioning, refer to the Maintenance
section of this manual for more instructions.
10.3.1 User Configuration
Change the Password
It is recommended that the password for 'Engineer' profile access should be changed from the default setting. The password is 8 digit
alphanumeric (case sensitive). A full 8 digit password should be used.
1. Select Device Setup.
2. Select Account.
3. Select Change Password.
4. Enter the password xxxxxxxx (eight characters, Alphanumeric, case sensitive) and press Enter.
5. The display shows a confirmation message, press OK.
6. Press Back to return to the Device Setup menu.
Note:
Some HART handheld devices contain a default password or hold a previously used password. Ensure that the desired password is
entered in before pressing Enter.
Set the Time and Date
Searchzone Sonik has a real time clock, which is factory set to UTC time.
Note:
Do NOT set HART to non-UTC time. The mobile App auto-translates detector's UTC timestamps to local timestamps. Changing
detector's UTC setting may cause connection issues.
1. Select Device Setup.
2. Select Account.
3. Select Set Time.
4. Enter the correct time (24 hour format) and press Enter.
5. The display shows a confirmation message, press OK.
6. Select Set Date.
7. Enter the correct date (dd/mm/yyyy format) and press Enter.
8. The display shows a confirmation message, press OK.
9. Press Back to return to the Device Setup menu.
10.3.2 Configure Warning, Inhibit and Over-range Signal Levels
The Warning, Inhibit and Over-range signal levels can be changed from the default values.
The default levels and ranges are:
Parameter
Fault
Inhibit
Warning
Over-range
The tolerance on the nominal current output value is ± 0.2 mA.
PN 2331M1220 Issue 1
Default
1 mA
2 mA
3 mA
21 mA
Searchzone Sonik − Technical Manual
10 Communications
Range
0 mA to 3.6 mA
1 mA to 3.6 mA
1 mA to 4 mA
20 mA to 22 mA
73
11/2018