Honeywell Sensepoint XCD Calibration Manual

Honeywell Sensepoint XCD Calibration Manual

Hide thumbs Also See for Sensepoint XCD:

Advertisement

Calibration Guide
Sensepoint XCD

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Sensepoint XCD and is the answer not in the manual?

Questions and answers

Summary of Contents for Honeywell Sensepoint XCD

  • Page 1 Calibration Guide Sensepoint XCD...
  • Page 3 flow rate of the test gas by a further 1 LPM, OR, to shelter the weather protection from the wind. 2. A full gas calibration of the sensor as described in the following section, using ONLY the Sensepoint XCD Gassing Cap (Part No.: S3KCAL).
  • Page 4 9. The display shows ‘SPAn’ with ‘YES’ flashing. 10. If span calibration is required use ‘’ proceed to the next step. If span calibration is not required, use ‘TS’ to select ‘No’ and ‘’ to return to configuration mode. Sensepoint XCD...
  • Page 5 13. Connect the regulator to the span gas cylinder. 14. Apply the span gas to the sensor using the Sensepoint XCD Gassing Cap. The live gas reading is displayed. When the reading is stable, use ‘’ to confirm span calibration.
  • Page 6 3. If a span calibration is to be performed, the span calibration gas should be applied to the sensor immediately after the zeroing procedure. Do not allow the sensor to return to ambient air conditions between steps 2 and 3. 4. Apply the span gas for two minutes before using ‘’ to confirm the span calibration. Sensepoint XCD...

Table of Contents