UV irradiation by an external signal (foot SW)
(when using one head)
When using a foot SW or other external signal to start the UV irradiation, use the [START]
input terminal for easier connection.
When using only one LED head, connect the head to [CH1], and connect the foot SW
between terminals [1] and [12].
Don't remove the shorting bar between terminals [11] and [12].
(Since the bar is used for an emergency stop signal, the UV irradiation is disabled if the
connection is off.)
Foot SW
UV irradiation
Preset UV
irradiation time
SW
5
Equipment
design
Connect the SW
between [START1]=[1]
and [COM]=[12]
by A-contact.