Replacement Of Parts - Emerson Rosemount Xstream Series Instruction Manual Addendum

Gas analyzer. flameproof variation for use in zone 1 and division 2 hazardous areas.
Hide thumbs Also See for Rosemount Xstream Series:
Table of Contents

Advertisement

X-STREAM XEFD

4.2 Replacement of Parts

Replacement of parts found defective is permitted only by using original
parts!
Violation voids the approvals and may cause explosions!
Parts beside the cast enclosure, the safety of the X-STREAM XEFD relies on:
Cable glands
NOT FOR analyzers for
vapor recovery applications!
Cable glands
suitable for all ATEX
approved variations
Cable entries stopping plugs
4-6
4.2
Replacement of Parts
EXPLOSION HAZARD BY REPLACEMENTS PARTS
Specifically for ATEX approved analyzers
Any ATEX approved flameproof type without
compound, suitable according EN 60079-14,
e. g.
- 8163/2 A2F M20 by Stahl
Any ATEX approved flameproof type with
compound, suitable according EN 60079-14,
e. g.
- ADE 1FC ISO 20 No. 05 by CAPRI,
- 8163/2–PXSS2K M20 by Stahl
Any ATEX approved flameproof type, e. g.
- 120209 by CAPRI
Male thread: M20 x 1,5
Spare part # ETC01754
Emerson Process Management GmbH & Co. OHG
Instruction Manual
HASXEDE-IM-EX
03/2012
or

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rosemount xstream xefd

Table of Contents