To switch differential mode on/off and enter the value:
Elcometer 206
1. Switch on the gauge.
2. Press the DIFF key.
The gauge will display DIFF OFF or DIFF
followed by a thickness value and flashing MM
(or IN), depending on which mode the gauge is
in.
3. Repeat step 2 to switch differential mode on or
off.
4. Assuming
DIFF
displayed, use the UP and DOWN arrow keys
to scroll to the desired nominal thickness
value.
5. Press the DIFF key once again to select the
nominal value entered.
24
value
flashing
MM
Elcometer 206DL
1. Switch on the gauge.
2. Press MODE key to activate features and
settings.
3. Press UP and DOWN arrow keys to scroll to
dIFF symbol.
The gauge will display DIFF OFF or DIFF
followed by a thickness value and flashing MM
(or IN), depending on which mode the gauge is
in.
is
4. Press SEND key to switch differential mode on
or off.
5. Assuming
DIFF
displayed, use the UP and DOWN arrow keys
to scroll to the desired nominal thickness
value.
6. Press SEND key to select the desired nominal
value and return to mode menu.
7. Press MODE key when finished.
value
flashing
MM
is