User Dialog Error Messages - X-Rite SP60 Series Getting Started

Sphere spectrophotometer
Table of Contents

Advertisement

Error Messages
Errors encountered during a measurement are displayed on the
instrument screen. All errors are accompanied by a long beep
and flashing yellow light. The error message is cleared from the
instrument screen by pressing the Enter key # .
Displayed Errors:
Measurement was
aborted by user
Calibration has
timed out
Calibration
required
Calibration has
failed
The battery is
getting low
Batteries are
dead
Batteries are
missing
Incorrect
Charger Voltage
The battery is
over–charged
Lamp is getting
weak, replace
soon
S P H E R E
Reason
Displays with an incomplete measurement or
calibration. Instrument was released too
soon.
The calibration interval time set in the
configuration has been reached. Calibration
is now required.
Displays anytime the instrument requires a
calibration.
Calibration failed. Make sure the instrument
is properly positioned on the reference.
This warning appears when the battery falls
below approximately 25% of full charge.
Measurements are still possible, but the
battery should be charged soon.
Displays when not enough battery power
remains to take measurement. The current
measurement is aborted.
The battery pack is not installed. Instrument
will not allow any measurements.
Wrong charger is connected or charger is
bad.
The battery pack is too hot. Remove battery
pack and let it cool off.
Reading lamp is at 50% strength or less from
its original intensity. Measurements are still
possible but the lamp should be replaced
soon.
S P 6 0 S E R I E S
S P E C T R O P H O T O M E T E R