Configuration With A Handheld Terminal - ABB LMT Series Operating Instruction

Magnetostrictive level transmitters foundation fieldbus
Hide thumbs Also See for LMT Series:
Table of Contents

Advertisement

L M T 1 0 0 A N D L M T 2 0 0 | M A G N E TO S T R I C T I V E L E V E L T R A N S M I T T E R S | F F O I/ L M T 1 0 0/ 2 0 0/ F F - E N R E V. A
7 Select Level Damping Time in the Easy Setup menu and press
Damping allows for smoothing of the device output step
response
8 Select Display 1 Line 1 View in the Easy Setup menu and press
This parameter sets the variable to be displayed on line 1 of the
product display
9 Set a tag in the Easy Setup menu and press
A tag is a quick way to identify the device

Configuration with a handheld terminal

The user can utilize a hand-held terminal to read out or
configure / calibrate the transmitter
Hand-held terminals such as Emerson Process 375, 475 and
AMS (provided the LMT Series EDD is downloaded and enabled
in the terminal)
Figure 26 Connection examples with Emerson 475
For additional information, refer to the operating instructions
included with the hand-held terminal
If the transmitter is configured in the factory, according to
customer specifications for the measuring point, all the user
needs to accomplish is the mounting and wiring of the
transmitter, as described The measuring point is now ready for
use
Each configuration step is subject to a plausibility check The
user can call up context-sensitive help at any time by pressing
the "F1" key Immediately after receiving the transmitter or
before changing the configuration, it is recommended to save
the existing configuration data to a separate data storage
media via the path: "File_Save"
27