4-wire connections

Connections for 4-wire local sensing are shown in Figure 4. Use 4-wire remote sensing for
the following source-measure conditions:
With 4-wire sensing selected, the 4W annunciator is on. The following menu sequence
selects 4-wire remote sensing:
Press CONFIG > select SOURCE V (or MEAS Ω) > select SENSE MODE > select 4-WIRE
NOTE
Figure 4
4-wire connections
Test circuit impedance is <1kΩ.
Optimum ohms, V-source, and/or V-measure accuracy is required.
Specified accuracies for both source and measure are achieved only using 4-wire
remote sensing.
4-WIRE
INPUT/
SENSE
OUTPUT
HI
LO
TERMINALS
ON/OFF
FRONT/
FRONT/
REAR
REAR
OUTPUT
SourceMeter Front Panel
Quick Results Guide
DUT
Sense Selection: 4-wire
7