First Use; Irrigation Solution; Irrigation Lines; Adjustments / Settings / Modes / Interface - Satelec IMPLANT CENTER 2 User Manual

Hide thumbs Also See for IMPLANT CENTER 2:
Table of Contents

Advertisement

- Check that the power switch (Fig. 2-5) is in
position O (OFF).
- Connect the power cord to the receptacl of the
device.
- Connect the power cord to a power socket with
a ground pin.
- Connect the control footswitch to the
footswitch connector (Fig. 2-2).
- Position the footswitch so that it is easily
accessible for your feet.
- Insert the brackets into the bracket holders
(Fig. 2-4).
- Connect the micro-motor cord to the connector
(Fig. 1-7).
- Align the I-S
URGE
connector electrical contacts, then screw the
motor onto the connector.
- Connect the N
EWTRON
2 handpiece cord to the connector (Fig. 1-9)
(according to option).
- Connect the N
EWTRON
cord (delivered according to option).
- Check that the device is close enough to the
work area to be used without pulling on the
cords. If it is not, move the device closer.
- Place the I-S
micro-motor on its rest.
URGE
- Place the N
EWTRON
the handpiece support.
- Suspend the bottles or bags of physiological
saline or sterile water from the brackets.
- Adjust the position of your device to suit your
angle of view.

4. 4 FIRST USE

Before using the I
MPLANT
time, all the equipment must be given routine
maintenance and/or sterilized according to the
procedures defined in chapter 11.
12
micro-motor and cord
scaler cord or the P
handpiece to the scaler
or P
2 handpiece on
IEZOTOME
C
2 for the first
ENTER
V - IRRIGATION SOLUTION
I
C
MPLANT
drugs. It must be used only with bottles or bags
of physiological saline or sterile water with an
individual capacity not exceeding one litre.
VI - IRRIGATION LINES
I
C
MPLANT
SATELEC irrigation lines.
VII - ADJUSTMENTS /
IEZOTOME
SETTINGS / MODES /
INTERFACE

7. 1 STARTUP

Important:
After stopping the device, wait four seconds
before restarting it.
Each time the device is switched on, the screen
displays program P1 of the I-S
displaying the welcome page.

7. 2 SETTINGS

The settings are saved when each program is
configured and are retrieved whenever that
program is selected.
2 is not designed to administer
ENTER
2 must be used exclusively with
ENTER
function after
URGE