7. 3 IRRIGATION FUNCTION

a) Flush / prime
The flush / prime function is available on the LCD
screen and on the footswitch.
Press button
(Fig. 3-1) or the footswitch
button (Fig. 11-2) (P
to switch on the flush / prime function.
The flush / prime function stays on as long as the
button is pressed.
The flush can be switched on while the device is
being used.
b) Flow rate adjustment
The irrigation flow rate is adjusted using buttons
and
(Fig. 3-4 and 5).
Each time the corresponding button is pressed,
the flow rate is adjusted.
The adjusted value is displayed on the LCD screen
(Fig. 3-2) and stored in the current program.
The flow rate can be adjusted while the device is
being used.
c) Irrigation ON/OFF
The irrigation is switched ON or OFF by pressing
button
(Fig. 3-3) or the footswitch button
(Fig. 11-4).
When irrigation is selected, the symbol
highlighted.
When irrigation is not selected, the symbol
is darkened.
and N
IEZOTOME
EWTRON

7. 4 LIGHT FUNCTION

The light function is displayed on the LCD screen.
Pressing button
function available for the N
functions ON or OFF.
modes)
The light function has a default light-off time-out
of nine seconds.
Switching off the light function becomes
effective when the time-out has elapsed.
If necessary, you can lengthen the time-out (refer
to 7.7).

7. 5 I-SURGE FUNCTION

a) Selecting the rotation direction
The rotation direction of the I-S
is controlled by the control footswitch button
(Fig. 11-2).
When clockwise rotation is selected, the symbol
is displayed on the LCD screen (Fig. 6-23).
When counterclockwise rotation is selected, an
audible signal is emitted and the symbol
displayed on the LCD screen (Fig. 6-23).
b) Program setting adjustment
Important:
When programming the maximum instrument
is
end speed, the entered value cannot exceed
the pre-programmed value.
Do not go below the minimum motor speed
value (100 rpm, 1:1 contra-angle).
Do not exceed the maximum pre-programmed
value of the instrument end torque, and do not
go below 10% of this value.
(Fig. 3-11) switches the light
and P
EWTRON
micro-motor
URGE
IEZOTOME
is
13