Satelec IMPLANT CENTER 2 User Manual page 18

Hide thumbs Also See for IMPLANT CENTER 2:
Table of Contents

Advertisement

The relative value of the audio volume is
displayed (Fig. 10-48).
b) Screen brightness adjustment
Adjust the brightness value
using buttons
and
The brightness is adjustable from 30% to 100%.
The brightness level setting is displayed (Fig. 10-
49).
c) Light off time-out adjustment
Adjust the time-out length value
by pressing buttons
47).
The length of the adjustable time-out (from nine
seconds to 18 seconds) is displayed (Fig. 10-50).
Contra-angle
Program
P1
P2
P3
P4
18
(Fig. 10-44)
(Fig. 10-46 and 47).
(Fig. 10-45)
and
(Fig. 10-46 and
Speed at
instrument
ratio
end
20:1
1200 rpm
20:1
800 rpm
20:1
15 rpm
20:1
30 rpm
d) Storing the settings
To store the modified settings, press button
(Fig. 10-52). The screen shown in Fig. 9
reappears.
e) Factory configuration reset
Press button
factory configuration settings.
The factory configurations for the P
N
functions are effective when button
EWTRON
(Fig. 10-52) is pressed. The screen shown in Fig.
9 reappears.
f) Software version
The version of the software is displayed at the
bottom of the screen (Fig. 10-51).
I
C
MPLANT
ENTER
the following settings:
I-S
mode:
URGE
Torque at
instrument
Irrigation
end
80 N.cm
80 ml/min.
80 N.cm
100 ml/min.
20 N.cm
100 ml/min.
20 N.cm
(Fig. 10-42) to restore the
2 is configured in the factory with
Functions
Implant site marking
Pilot drilling
Boring / tapping
0
Screwing
and
IEZOTOME

Advertisement