e) Sterilization
The tips must be sterilized individually in an
autoclave,
according
parameters:
- autoclave: type B, compliant with standard EN
13060;
- sterilization temperature: 134°C;
- sterilization stabilization time: 18 minutes;
- pressure: 2 Bars min.
f) Storage
Store the sterilized products in a dry place
protected from dust. Before re-use, if the
integrity of the packaging is not certain,
repackage then re-sterilize in accordance with
the defined protocol.
If contamination is visible inside the bag, destroy
the product.
g) Disposal of the product
Dispose of the product in receptacles for waste
materials of healthcare activities involving
infectious risks.
CAUTION: Not all autoclaves can reach 134°C.
Not all autoclaves draw a pre-vacuum. Please
refer to your autoclave manufacturer for
specific sterilization instructions.
XII - MONITORING /
MAINTENANCE
Important:
In the case of a fault, you are advised to
contact the supplier of your device. Do not use
unauthorized repairers, who might make your
device dangerous for you and your patients.
to
the
following
12. 1 MONITORING
Regular monitoring of the device and its
accessories is necessary in order to detect any
insulation fault or damage. Replace them if
necessary.
It is important to keep the control unit
ventilation vents clean in order to avoid
abnormal heating.
12. 2 MAINTENANCE
I-S
micro-motor
URGE
SATELEC recommends inspection or overhaul of
the I-S
micro-motor at least once a year.
URGE
12. 3 FUSE REPLACEMENT
I
C
2 is protected by two fuses located
MPLANT
ENTER
in the power receptacle (Fig. 2-1).
The replacement procedure is as follows:
- Switch the device OFF (position O).
- Unplug the power cord from the electricity
supply.
- Unplug the power cord from the receptacle
(Fig. 2-1). Insert the tip of a flat screwdriver
into the notch above the fuse drawer to release
it.
- Remove the old fuses.
- Replace the old fuses with fuses of the same
type and rating.
- Push the fuse drawer back into its housing until
it clicks to indicate the correct position.
- Plug the power cord into the receptacle (Fig. 2-
1). Plug the power cord into the electrical
supply.
inspection/overhaul:
25