Summary of Contents for Satelec Suprasson P5 Booster

  • Page 1 Suprasson P5 Booster ® User's manual • Notice d'utilisation • Manual de uso Benutzerhandbuch • Manuale d'uso • Handleiding voor de gebruiker...
  • Page 3 • T I NG VI T Ư NG D N S D NG • F ’ RANÇAIS ANUEL DE L UTILISATEUR • E SPAÑOL ANUAL DE USO • D EUTSCH ENUTZERHANDBUCH • I ’ TALIANO ANUALE D • N EDERLANDS ANDLEIDING VOOR DE GEBRUIKER...
  • Page 4 ENGLISH ENGLISH...
  • Page 5: Table Of Contents

    ONTENTS I - PRESENTATION AND INSTALLATION Presentation Technical description Installation II - MAINTENANCE AND STERILIZATION Suprasson handpiece maintenance Scaling tip maintenance Water filter maintenance and replacement Device maintenance Disposal and recycling Trouble shooting III - LIABILITY IV - SYMBOLS The list of tips included in this manual is not exhaustive. Contact your local dealer for more information.
  • Page 6: I - Presentation And Installation

    I - PRESENTATION AND INSTALLATION 1. 1 PRESENTATION SATELEC's Suprasson P5 B is an ultrasonic OOSTER generator for the main dental applications : scaling and endodontic treatment. The Suprasson P5 B is comprised of the OOSTER following elements : • 1 green flexible tube (2 m long) with integral particle filter for connection onto the mains water supply.
  • Page 7: Installation

    Note: before using for the first time, clean, See the TipBook for the power settings for decontaminate and sterilize the accessories. SATELEC tips. Before using the unit for the first time you are recommended to plug in the accessories and...
  • Page 8: Maintenance And Sterilization

    II - MAINTENANCE AND Always be sure to replace tips subject to deterioration due to wear or accidental impact STERILIZATION (fall, etc.). b) Sterilization of tips Tips should be cleaned before sterilization The following instructions for maintenance using a cotton swab and alcohol, wipes or by and/or sterilization must be carried out before immersion in an ultrasonic bath.
  • Page 9: Device Maintenance

    Disposal of the equipment and its accessories should be carried out by a collection centre for electronic equipment or by returning them to SATELEC's After Sales Service. In no case should the device and its accessories be disposed of by the user.
  • Page 10 • Presence of fluid or moisture between • Dry any moisture, handpiece and cord especially on electrical contacts • Blown fuse • Return to SATELEC After Sales Service No spray • Defective water tube connection • Check water supply • No water pressure •...
  • Page 11: Liability

    Note without prior notice. All the information necessary to repair the parts of the device specified as repairable by SATELEC Environment is available on request to technical staff of the For patients and/or users fitted with a pace- network of SATELEC approved dealers.
  • Page 12: Symbols

    230 V H2O : 1 - 5 bars Class II device Service / opération : Intermittent 5 mn/10 mn S/N : Made in FRANCE by SATELEC ZI du Phare 33700 MERIGNAC FRANCE BF Type Caution, refer to accompanying documents Functionnal earth terminal...