Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These specifications may not be applicable in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and persons.
(a): The electromagnetic field strengths of fixed radiofrequency emitters, such as base stations for
mobile telephones (cellular/cordless), mobile radios, amateur radio, AM/FM radio broadcasts and
TV broadcasts cannot be determined exactly by theory. To assess the electromagnetic
environment due to fixed radiofrequency emitters, an electromagnetic environment
measurement must be made. If the measured radiofrequency field strength in the immediate
environment where the product is used exceeds the compliance level specified above, the
performance of the product must be tested to verify whether it conforms to the specification. If
abnormal performance is observed, additional measures may be necessary, such as reorienting or
relocating the product.
(b): In the 150 kHz to 80 MHz frequency range, the electromagnetic field strengths must be less than
3 V/m.
13. 4 RECOMMENDED SEPARATION DISTANCES
I
C
2 is intended for use in an electromagnetic environment in which radiated
MPLANT
ENTER
radiofrequency disturbances are controlled.
The I
C
MPLANT
ENTER
maintaining a minimum distance between portable and mobile radiofrequency communications
equipment emitters and the I
equipment, as recommended in the table below.
Rated max. power
of the emitter (W)
0,01
0,1
1
10
100
For emitters rated at max. power not listed above, the recommended separation distance d in meteres (m) can be estimated
using the equation applicable to the frequency of the emitter, where P is the max. power rating of the emitter in watts (W)
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These specifications may not be applicable in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and persons.
30
2 user and/or installer can help prevent electromagnetic interference by
C
MPLANT
ENTER
Separation distance in metres (m) according to emitter frequency
150 KHz to 80 MHz
d = 1.2
0.12 m
0.38 m
1.2 m
3.8 m
12 m
according the manufacturer.
2, according to the maximum output power of the
80 MHz to 800 MHz
d = 1.2
0.12 m
0.38 m
1.2 m
3.8 m
12 m
800 MHz to 2,5 GHz
d = 2.3
0.23 m
0.73 m
2.3 m
7.3 m
23 m