Adjusting The Offset Parameters - Trotec Speedy C12 Service Manual

Individual m ar king system
Table of Contents

Advertisement

Adjusting the Offset parameters:

The Speedy uses so called offset parameters (X and Y) to make sure that the zero
position of the software is exactly in the crosspoint of the rulers. This is important for
precise positioning of jobs.
Additionally we use the Z offset to finetune the automatic focus. (only for focussing
using the software)
An offset is a certain distance from the home position that is given by the limit
switches to the edge of the rulers (for X and Y) or the focus level of the lens (Z).
Whenever you see that the position of your engraving on the table is not the same as
in your desktop publishing program you have to adjust the offsets.
Steps for adjusting the X and Y offsets:
1. Make sure that the ruler angle has been adjusted correctly.
2. Focus your beam on the surface of the rulers and glue some paper tape in the
crosspoint.
3. Move the engraving head to 0/0 (Engraver>move laser)
4. Fire the test pulse until you can see a small spot in the paper tape.
NORMAL.DOT
max
Tab „Offset" in the Service menu
checking the zero position
Gespeichert in:G:\Support\Dokumentationen_CDs\8003Speedy\Service Manual\Servicemanual 8003.doc
I N D I V I D U A L M A R K I N G S Y S T E M S
Y
X
Gespeichert am:2003/03/10
Version 14
Seite 31 von 87