Process seal antenna
Hide thumbs Also See for Rosemount 5408:

Advertisement

Quick Start Guide
Step 9: Attach and tighten the covers
1. Verify the cover jam screws are completely threaded into the housing.
2. Attach and tighten the covers. Make sure the covers are fully engaged.
3. Turn the jam screw counterclockwise until it contacts the cover.
Required for explosion-proof/flameproof installations only.
4. Turn the jam screw an additional ½ turn counterclockwise to secure the
cover.
Step 10: Connect the power supply
It may take up to 15 seconds before the LCD display lights up.
16
Cover jam screw
(one per side)
October 2017