Bed Pan Flusher / Sanitizer - Energy Saver
With Hands Free Operation
Operation, Maintenance and Installation Manual
Plumbing
Soil Line
There is no high temperature water discharged from this machine so no special high temperature
pipework is necessary.
Soil line connection is by a pan collar or other preferred method.
If the belled end on the
polyethylene moulded trap is not required it can easily be cut off to provide a straight pipe
connection.
Level the unit by using the flanged screw in legs and if possible maintain approximately 100mm
floor clearance for ease of floor cleaning. We recommend some of the leg flanges are fixed to the
floor via stainless self tapping screws, dyna bolts or similar to prevent sideways movements and
damage to services and soil line connections.
Soil line to protrude from the floor or wall at a minimum of 100mm.
Water
ONLY COLD WATER CONNECTIONS IS REQUIRED. The machine can be connected to any
normal mains pressure cold water supply as an AIRBREAK is incorporated in the design of the
cold water tank.
The water supply should be connected to an isolating valve or cistern stopcock placed
approximately 1200mm from the floor to the left-hand side of the machine. (Preferably not behind
the machine).
Minimum water pressure 95 Kpa.
rd
31
Issue 24
Page
23
August 2013