Bed Pan Flusher / Sanitizer - Energy Saver
With Hands Free Operation
Operation, Maintenance and Installation Manual
Design Parameters
The Malmet Sanitizer has been designed within the following parameters:
a)
A single bedpan with lid and two urine bottles can be emptied, cleaned and sanitized during
each automatic cycle.
The utensils that can be cleaned in the machine are:-
i)
Standard size bed pans
ii)
Standard commode bowl
iii)
Standard male and female urine bottles
iv)
Most plastic urine bottles including male non-spill and female cervix
v)
Small Slipper pan
b)
The cradle is designed to ensure that utensils are not dislodged during the cleaning cycle,
the contents are emptied during door closure.
c)
The chamber and door are self-cleaning and do not permit water or soil to remain after a
properly completed cycle. Steam sanitation ensures all internal surfaces are totally clean
and safe.
d)
The flush and clean stage:
i)
Removes the soil
ii)
Clears the trap
e)
A complete cycle is completed in approximately four (4) minutes for the 20 Amp Sanitizer
and six (6) minutes and thirty (30) seconds for the 10 Amp Sanitiser.
NOTE: Dependant on ambient air temperature.
Note: If the machine is turned off or on initial start up, the first cycle will take longer as the
machine has to raise the temperature of cold water to 90° C.
Disconnect power when servicing
Issue 24
5
Page
rd
23
August 2013