Malmet Energy Saver ES910 Operation, Maintenance And Installation Manual page 19

Bed pan flusher/ sanitizer
Hide thumbs Also See for Energy Saver ES910:
Table of Contents

Advertisement

Bed Pan Flusher / Sanitizer - Energy Saver
With Hands Free Operation
Operation, Maintenance and Installation Manual
Fault 06
Hot Water Tank Fill Time Exceeded
(high level not reached during pre-fill, resulting in time out)
Cause
Fault 07
Hot Water Tank Low Water Level Sensed During Cycle
Cause
Fault 08
Door Closure Fault – Not Confirmed Door Open
(door switch still indicates closed)
Cause
Fault 09
Failed To Reach Sanitize Start Temperature Resulting In Time Out
Cause
Fault 10
Sanitize Temperature Not Maintained Resulting In Cycle Time Out
Cause
Fault 12
Hot Tank Short Temperature Sensor
Fault 13
Hot Tank Open Circuit Temperature Sensor
Issue 24
Faulty fill solenoid
Faulty water level probe
Cold water tank empty or not filling quick enough
Low water pressure
Faulty water level probe
Cold water tank empty or not filling quick enough
Leaking tank or hose
Element stuck on
Condensing coil blocked
Door limit switch stuck on
Steam leak
Final rinse solenoid stuck open
Faulty temperature sensor hot water tank
Blown element leg
Steam condensate leaking cold water into chamber
Faulty temperature sensor chamber
Door not reaching interlocking micro switch. Door not fully closed.
Over Temperature Board (Power to element open circuit)
Element overtemp protection manual reset thermal cut-out switch
open circuit.
Steam leak
Final rinse solenoid stuck open
Faulty temperature sensor hot water tank
Blown element leg
Steam condensate leaking cold water into chamber
Fault temperature sensor – chamber
Page
19
rd
23
August 2013

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for Malmet Energy Saver ES910

This manual is also suitable for:

Energy saver es930

Table of Contents