Plumbing; Technical Specifications - Malmet Energy Saver ES910 Operation, Maintenance And Installation Manual

Bedpan washer disinfection. detergent model with hands free operation
Hide thumbs Also See for Energy Saver ES910:
Table of Contents

Advertisement

Bedpan Washer Disinfector - Energy Saver
Detergent Model With Hands Free Operation
Operation, Maintenance and Installation Manual

Plumbing

Soil Line
There is no high temperature water discharged from this machine so no special high temperature pipework
is necessary.
Soil line connection is by a pan collar or other preferred method. If the bell end on the polyethylene
moulded trap is not required it can easily be cut off to provide a straight pipe connection.
Level the unit by using the flanged screw in legs and if possible maintain approximately 100mm floor
clearance for ease of floor cleaning. We recommend some of the leg flanges are fixed to the floor via
stainless self tapping screws, dyna bolts or similar to prevent sideways movements and damage to services
and soil line connections.
Soil line to protrude from the floor or wall at a minimum of 100mm.
Water
ONLY COLD WATER CONNECTIONS IS REQUIRED. The machine can be connected to any normal mains
pressure cold water supply as an AIRBREAK is incorporated in the design of the cold water tank.
The water supply should be connected to an isolating valve or cistern stopcock placed approximately
1200mm from the floor to the left-had side of the machine. (Preferably not behind the machine)
Minimum water pressure 95 Kpa.
MODEL
ES 910 20A DET
ES930 10A DET
NOTE: Values may change due to operating and supply service conditions.
Issue 21

Technical Specifications

POWER AND WATER CONSUMPTION DATA
Avg Cycles per/Hr
Avg Cycle min/sec
11
7.7
Avg kWh per cycle
5.3
7.5
Page 33
CW Avg Lt per cycle
0.300
36
0.236
35.5
14 September 2017