Cleaning; Disposal And Recycling; Guide De Dépannage - Kinetec Manosplint Dry Heat Oven User Manual

Table of Contents

Advertisement

Cleaning

Before any intervention, switch OFF the unit by disconnecting the power cord.
Before cleaning the heating surfaces, make sure that the appliance is cold.
Use a disinfectant (alcohol-free solution or <5% alcohol) in wipes or spray.
If in doubt about the compatibility of cleaning agents with the parts of the device or with the
material inside, contact your Kinetec® distributor.
Notes:
-
Residues of thermoplastic material may remain stuck on the non-stick surfaces.
-
Once the device is cold, the residues must be removed with a dry cloth.
-
We recommend cleaning the heating surfaces daily.
Do not scrape the heating surfaces with pointed or sharped objects, risk of irreversible damage to the non-
stick surface.

Disposal and recycling

Packaging: The packaging must be separated into plastic and paper / cardboard components and taken to
special recycling sites.
Device: It contains electronic components, cables, aluminium, steel and plastic parts. When the machine is no
longer operational, disassemble it, separate it into different types of material and take these to authorized
recycling centers or return the machine to Kinetec SAS for destruction. Or contact the local authorities to
determine the appropriate method of disposal for parts and accessories that are potentially hazardous to the
environment.
Guide de dépannage
After connecting the power cord to an electrical socket and after turning on the power:
The ON/OFF switch (7) and/or the display are not lit:
-
Check that the electrical socket is live using another device or a voltmeter.
-
Replace the fuse(s) of the IEC connector (5) with fuses of the same type and calibre: see "Technical
specifications".
-
Contact your Kinetec® specialist.
The non-stick surfaces do not heat, the current temperature indication on the display does not change:
-
Check that the top cover is closed and locked.
-
If the problem persists, the top cover locking system or the top cover opening detection system is out of
adjustment. Contact your Kinetec® specialist and ask for the procedure for adjusting the closing systems
of the top cover.
Cannot set the target temperature below 50°C (122°F) or above 90°C (194°F):
-
For safety reason, the setting of the target temperature is limited between 50°C and 90°C (122°F/194°F).
EN
11/12

Advertisement

Table of Contents
loading

Table of Contents