Maintenance; Troubleshooting Guide; Cleaning; Disposal And Recycling - Kinetec Prima XL User Manual

Knee continuous passive motion machine
Table of Contents

Advertisement

Maintenance

After 2,000 hours of operation, or once a year, the device Kinetec Prima XL™ requires lubrication and maintenance
operations (lubrication of the joints, pointer stops and ball screws).
When the machine is switched on the LED blinks 5 times to indicate that the service interval has been reached.
Despite this warning, you can continue to use your Kinetec Prima XL™ by pressing the hand control switch, but you
should contact your nearest KINETEC
possible.
An after-sales service inspection sheet and the technical catalogue are available on request from your KINETEC
distributor.
WARNING
Before using this machine, always check that the electrical socket is in good condition and is
:
suitable for the splint power supply cord. Only use the original cable supplied with the machine.
Check that the cables remain free around the device so that they do not get damaged.
WARNING
Before using this machine, always check that the machine is not damaged, in particular the
:
protective housings.
When the machine is no longer in working condition, please return it to us, together with its accessories, for
destruction.

Troubleshooting guide

A spare parts list and technical catalogue are available on request from your KINETEC
If, after connecting the power supply cable to the power supply and switching
on the machine:
• The LED on the control panel does not light up:
- Check that the electrical socket is live using another device or voltmeter.
- Replace the fuse(s) (20) of the connector with fuses of the same type and
calibre:
2 fuses T 750 mA 250V (6.3 x 32) (KINETEC
- If the display still does not indicate any information, contact your nearest
®
KINETEC
technician.
• Your machine does not work but the LED on the control panel lights up,
Press the hand control switch once.
If your machine still does not work, contact your nearest KINETEC
• The LED on the control panel indicates a fault:
The LED blinks 1 time  angle measurement function failure (copy potentiometer problem)
The LED blinks 2 times  no movement
The LED blinks 3 times  abnormal motor consumption or motor short-circuit
The LED blinks 4 times  the motor PCB is providing power but motor consumption is zero.
The LED blinks 5 times  Service Time ≥ 2,000 hrs, maintenance required
The LED blinks 6 times  not enough power
The LED blinks 7 times  START/STOP switch failure
Contact your nearest KINETEC

Cleaning

Before carrying out any cleaning operation, SWITCH OFF the unit and disconnect the power supply. In order to ensure
optimal hygiene, you are advised to clean the machine for each new patient.
Cleaning should be carried out in the environmental conditions specified in the
"Technical Specifications" section below.
Use a DISINFECTANT product (alcohol-free or <5% alcohol solution) in spray.
In order to ensure optimal hygiene, you are advised to clean the covers for each new patient. All the consumables
enable hazard-free disposal.

Disposal and recycling

a • Packaging: The packaging must be separated into plastic and paper / cardboard components and taken to
special recycling sites.
b • KINETEC™ patient pad kit: Clean with a disinfectant product then take it to special recycling sites.
c • Unit: It contains electronic components, cables, aluminium, steel and plastic parts. When the machine is no longer
operational, disassemble it, separate it into different types of material and take these to authorised recycling
centres or return the machine to Kinetec for destruction. Or contact the local authorities to determine the
appropriate method of disposal for parts and accessories that are potentially hazardous to the environment.
Kinetec Prima XL™
®
technician to have the maintenance operations carried out as soon as
®
order: 4610007434).
®
technician.
EN
®
distributor.
20
®
technician.
®
8/11