Selective Mobilization Of The Dip Joint; Product Information; Maintenance; Cleaning - Kinetec Maestra Manual

Hide thumbs Also See for Maestra:
Table of Contents

Advertisement

Possibles movements

Selective mobilization of the DIP joint

The KINETEC Maestra provides a motion from 0° to 70°.
• Choice of the Selective mobilization of the DIP joint movement:
- Press 3 times
- Or select a program

Product information

Maintenance

After 2,000 hours of operation, KINETEC Maestra requires a few lubrication and maintenance operations (lubrication of the joints, pointer stops and ball screws).
The need for maintenance is indicated by display of the message SERV. MOTOR when the system is switched on.
Despite that warning, you can continue to use your KINETEC Maestra by pressing START, but you should contact your nearest KINETEC technician to have the
maintenance
operations conducted as soon as possible.
• A motor running time counter is available. Simultaneously press keys
RESET TIME 924H
Reset: limit low

Cleaning

Before conducting any cleaning operation, SWITCH the unit OFF and disconnect the power supply.
Use a DISINFECTANT (PROPANOL/ISOPROPANOL or ALDEHYDE-based solution). Spray the disinfectant on the SURFACES (plastic shells and metal components).
In order to ensure optimal hygiene, you are advised to clean the covers for each new patient. All the consumables enable hazard-free disposal.

Technical specifications

Product:
Weight: 8 Kg (without case)
Unit dimensions:
35 x 41 x 30cm
14 x 16 x 12 inches
Angular limits: see page 6
Spedd: from 150 to 440°/min
Patient sizing:
all hand sizing until 7 years old
GB
18
and adjust your parameters (see page 9).
(see page 13).
this counter can be resetting by pressing the key
Electricity:
Power supply: 100-240 V~
Frequency: 50-60 Hz
Power consumption: 50 VA
Device of type B class I
IP 20.
Fuse T 750mA 250V 6.3x32mm
(KINETEC order: 4610007434)
and
, the displays shows
.
Environment
- Storage/transport conditions:
Temperature: -40 to 70°C / -40 to 160°F
Relative humidity: up to 90%
- Operating conditions:
Room temperature: 10 to 40°C / 50 to 105°F
Relative humidity: up to 80%