Maintenance; Troubleshooting Guide; Cleaning; Disposal And Recycling - Kinetec Spectra Essential User Manual

Passive knee mobilisation device
Hide thumbs Also See for Spectra Essential:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 17

Maintenance

After 2,000 hours of operation, or once a year, the device Kinetec Spectra Essential™ requires lubrication and
maintenance operations (lubrication of the joints, pointer stops and ball screws). The need for maintenance is
indicated by display of the message SERVICE TIME Mx when the system is switched on.
Despite this indication, you can continue to use your machine by pressing [START], but you should contact your
nearest Kinetec
technician to have the maintenance operations carried out as soon as possible.
®
An after-sales service inspection sheet and the technical catalogue are available on request from your Kinetec
distributor.
WARNING
:
Before using this machine, always check that the electrical socket is in good condition and is
suitable for the splint power supply cord. Only use the original cable supplied with the machine.
Check that the cables remain free around the device so that they do not get damaged.
WARNING
:
Before using this machine, always check that the machine is not damaged, in particular the
protective housings.
When the machine is no longer in working condition, please return it to us, together with its accessories, for
destruction.
A motor running time counter is available by simultaneously pressing keys
the displays shows RESET TIME 215H (this is an example).
This counter can be reset by pressing the key

Troubleshooting guide

A spare parts list and technical catalogue are available on request from your Kinetec
If, after connecting the power supply cable to the power supply and switching
on the machine:
• The display does not indicate any information:
- Check that the electrical socket is live using another device or voltmeter.
- Replace the fuse(s) (25) of the connector with fuses of the same type and
calibre: 2 fuses T 750 mA 250V (6.3 x 32) (Kinetec
- Check that the hand control is connected properly.
- If the display still does not indicate any information, contact your nearest
Kinetec
technician.
®
• Your machine does not work and the display indicates 50 STOP 25 115,
Press [START] again.
If your machine still does not work, contact your nearest Kinetec
• Your machine does not work and the display indicates:
"SERVICE D1": angle measurement function failure,
or "SERVICE D2": no movement,
or "SERVICE D3": abnormal consumption,
or "PUSH STOP/START": power failure or disconnected motor.
switch the machine off then on,
contact your nearest Kinetec

Cleaning

Before carrying out any cleaning operation, SWITCH OFF the unit and disconnect the power supply. In order to ensure
optimal hygiene, you are advised to clean the machine for each new patient.
Cleaning should be carried out in the environmental conditions specified in the
"Technical Specifications" section below.
Use a DISINFECTANT product (alcohol-free or <5% alcohol solution) in spray.
In order to ensure optimal hygiene, you are advised to clean the covers for each new patient. All the consumables
enable hazard-free disposal.

Disposal and recycling

a • Packaging: The packaging must be separated into plastic and paper / cardboard components and taken to
special recycling sites.
b • Kinetec
patient pad kit: Clean with a disinfectant product then take it to special recycling sites.
®
c • Unit: It contains electronic components, cables, aluminium, steel and plastic parts. When the machine is no longer
operational, disassemble it, separate it into different types of material and take these to authorised recycling
centres or return the machine to Kinetec SAS for destruction. Or contact the local authorities to determine the
appropriate method of disposal for parts and accessories that are potentially hazardous to the environment.
Kinetec Spectra Essential™
.
order: 4610007434).
®
technician if the same message is displayed
®
EN
and
distributor.
®
25
technician.
®
®
;
9/12

Advertisement

Table of Contents
loading

Table of Contents