15. Product information

15.1.

Maintenance

After 2,000 hours of operation, or once a year, the device Kinetec Maestra™ requires lubrication and
maintenance operations (lubrication of the joints and pointer stops). When the machine is switched on the
message SERVICE TIME Mx is displayed to indicate that the service interval has been reached.
Despite this indication, you can continue to use your machine by pressing [START], but you should contact your
nearest KINETEC
technician to have the maintenance operations carried out as soon as possible.
®
An after-sales service inspection sheet and the technical catalogue are available on request from your
KINETEC
®
distributor.
WARNING
Before using this machine, always check that the electrical socket is in good condition
:
and is suitable for the splint power supply cord. Only use the original cable supplied with
the machine. Check that the cables remain free around the device so that they do not
get damaged.
WARNING
:
Before using this machine, always check that the machine is not damaged, in particular
the protective housings.
When the machine is no longer in working condition, please return it to us, together with its accessories, for
destruction.
A motor running time counter is available by simultaneously pressing keys
the displays shows RESET TIME 215H (this is an example).
This counter can be reset by pressing the key
15.2.

Troubleshooting guide

A spare parts list and technical catalogue are available on request from your
KINETEC
distributor.
®
If, after connecting the power supply cable to the power supply and
switching on the machine:
- Check that the electrical socket is live using
another device or voltmeter.
- Replace the fuse(s) (1) of the connector with fuses of the same type and
calibre: 2 fuses T 750 mA 250V (6.3 x 32) (KINETEC
- Check that the hand control is connected properly.
- If the display still does not indicate any information,
contact your nearest KINETEC
press [START] again.
If your machine still does not work, contact your nearest KINETEC
"SERVICE D1": angle measurement function failure,
or "SERVICE D2": no movement,
or "SERVICE D3": abnormal electrical consumption by motor,
switch the machine off then on,
contact your nearest KINETEC
15.3.

Cleaning

to ensure optimal hygiene, you are advised to clean the machine for each new patient.
Cleaning should be carried out in the environmental conditions specified in the
"Technical Specifications" section below.
Use a DISINFECTANT product (alcohol-free or <5% alcohol solution) in spray (plastic cases and metal
components).
In order to ensure optimal hygiene, you are advised to clean the covers for each new patient. All the
consumables enable hazard-free disposal.
15.4.

Disposal and recycling

a
Packaging: The packaging must be separated into plastic and paper / cardboard components and
taken to special recycling sites.
KINETEC
patient pad kit: Clean with a disinfectant product then take it to special recycling sites.
®
It contains electronic components, cables, aluminium, steel and plastic parts. When the machine is
no longer operational, disassemble it, separate it into different types of material and take these to authorised
recycling centres or return the machine to Kinetec SAS for destruction. Or contact the local authorities to
determine the appropriate method of disposal for parts and accessories that are potentially hazardous to
the environment.
Kinetec Maestra™
.
order: 4610007434).
®
technician.
®
®
technician if the same message is displayed.
EN
and
50 STOP 25 115,
technician.
®
;
1
22 / 26

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents