 DISABLE turns off the relay and removes it's icon from the HOME screen
Setting up an Alarm Relay
 Choose ALARM
 Press CHANGE to enter new values
 Choose ON Set Point, Press OK
 Enter value using ▲ or ▼and NEXT to move to the next digit, press OK, press BACK (Min –Max values
indicate the range of acceptable values)
 Choose Expiration, Press OK,
 Choose time from drop down menu using ▲ or ▼, press OK, press BACK
 Choose OFF Set Point, Press OK
 Enter value using ▲ or ▼and NEXT to move to the next digit, press OK, press BACK
 Choose Delay ON, Press OK
 Enter value using ▲ or ▼and NEXT to move to the next digit, press OK, press
BACK
 Choose Delay OFF, Press OK
 Enter value using ▲ or ▼and NEXT to move to the next digit, press OK, and
press BACK when done to exit Relay 1.
 Repeat for Relay 2 and Relay 3.
3.3.4 E
M
XIT
ENUS AND
 Press HOME Key to return to the Home Menu Screen
 Press Hold to turn OFF Hold
 Press EXIT Key to exit the menu
 "Save Changes?" press YES
 Choose Display Mode, DATA, mV or GRAF by pressing selection Key. The
selection key displays which screen will be displayed next.
o The type of graphical display used, Line, Bar or Gauge is selected in
CONFIG → XMTR → LCD → GRAPH → LINE, GAUGE, BAR
3.3.5 S
S
U
ENSOR
TART
All sensors are supplied with protective caps over the sensing end. Remove the cap(s) from the sensor before
installing in the process. All sensors were calibrated at the factory before shipment, no calibration should be
necessary before use. Allow the sensor to equilibrate to the process solution conditions for ½ hour before
verifying the reading against a grab sample. If calibration is required follow the instruction in Section 4.0 below.

3.4 USER SELECTABLE OPTIONS

3.4.1 S
L
CREEN
IGHTING
LED back lighting is available on AC and DC powered instruments only.
Contrast can be adjusted for optimal viewing. The Backlight can be adjusted to timeout after a set period of time
or remain on.
Location: CONFIG → XMTR → LCD → Set Up → CONT, BACK LIGHT