1.1 FEATURES

Panel Mounted System, Easy Installation
Plumb and Play Design, Ready to Use
Automatic pH Compensation, No Expensive Reagents to
mix or spill
Automatic Flow Control, Eliminates Pressure Regulators
and Rotameters
T80 Transmitter Capability, Dual Measurements,24VDC or
110/220 VAC Power, Graphical Plots
Compliant with EPA Method 334.0

1.2 SPECIFICATIONS

1.2.1 S
F
ENSORS AND
Chlorine Sensor:
Digital protocol, Potentiostatic, Gold cathode/Silver-Silver Halide anode, 316L SS counter electrode
pH Sensor:
Digital S80 protocol, 316L stainless steel body with replaceable electrode cartridge
Measurement Range:
Chlorine: 0.05 to 20 ppm (High Range)
0.005 to 2.00 ppm (Low Range)
pH: 4 to 14 pH
Operating Temperature:
0° C to 50° C (32° F to 122° F)
Min/Max Flow:
38 L/hr. to 300 L/hr. (10 gal/hr. to 80 gal/hr.)
Wetted Materials:
PVC, PP, PVDF, PTFE, Glass, 316 SS
Process Connections:
Input ¼" FNPT with barb fitting, Drain ¾" FNPT
Response Time:
T90 in 2 minutes
Electrolyte Life:
3 to 6 months
1.2.2 TC80 A
NALYZER
Measurements:
Chlorine: 0.00 ppb to 22.00 ppm
pH: 0.00 to 14.00 pH
pH Compensation of Total Chlorine:
pH 4 - 12
Display:
128 x 64 pixels (2.75" x 1.5") LCD, Black on Grey background, Blue on White background with LED backlight on
Outputs:
(1) 4-20 mA for Total Chlorine set to Sensors Range