ECD H10C Instruction Manual

Microprocessor based conductivity, tds, salinity, temperature hand held meter

Advertisement

Quick Links

Model H10C
Instruction Manual
Microprocessor based
Conductivity, TDS, Salinity, Temperature
Hand Held Meter
Electro-Chemical Devices, Inc.
1500 N. Kellogg Dr.
Anaheim, CA 92807 USA
Telephone: +1-714-695-0051
www.ecdanalytical.com

Advertisement

Table of Contents
loading

Summary of Contents for ECD H10C

  • Page 1 Model H10C Instruction Manual Microprocessor based Conductivity, TDS, Salinity, Temperature Hand Held Meter Electro-Chemical Devices, Inc. 1500 N. Kellogg Dr. Anaheim, CA 92807 USA Telephone: +1-714-695-0051 www.ecdanalytical.com...
  • Page 2: Table Of Contents

    Contents General Introduction ............... 2 Initial Inspection ..............2 Waterproof ................2 Installing the Batteries ............. 3 Displays and Key Functions ............4 Keys ..................5 OPERATIONAL PROCEDURES ........... 6 A. Preparing Standard Solutions ......... 6 B. Calibration ..............6 TDS .................
  • Page 3: General Introduction

    TDS simultaneously with user prompts and mode indicators. The unit prompts the user through calibrations and measurement procedures. The Model H10C is available with a single four wire conductivity cell that measures from 2 µS to 200 mS. Its features include automatic conductivity ranging, automatic temperature compensation, long battery life and 50/60 Hz AC noise rejection.
  • Page 4: Installing The Batteries

    Installing the Batteries The Model H10C meter is packaged with 4 “AAA” alkaline batteries required for operation. To insert the batteries into the meter, follow the procedure outlined below. Use a screw driver to remove the two screw and battery cover to expose the battery compartment.
  • Page 5: Displays And Key Functions

    Displays and Key Functions 1. CONDUCTIVITY- 8. CELL- Displays when measuring conductivity Indicates conductivity cell constant value 9. °C/%- Displays during calibration: 2. BAT- ℃: Indicates temperature reference unit. Low Battery Indicator %: Indicates temperature coefficient unit. 3. UNCOMPENSATED- 10. ppt- Distinguishes between temperature...
  • Page 6: Keys

    Keys On/Off- Powers on and shuts off the meter. Mode- Selects display mode. In normal operation, press Mode to sequentially display compensated conductivity, salinity, total dissolved solids (TDS) and uncompensated conductivity. In calibration mode, this key exits the current calibration and displays the next calibration parameter. Up/Down- Increases or decreases the displayed value as desired Cal-...
  • Page 7: Operational Procedures

    Here are some standard solutions the user can prepare to calibrate the probe of the model H10C. Standard solution of 1413uS at 77°F (25°C): Accurately weight out 0.746 grams of research grade dried Potassium Chloride (KCL). Dissolve in 1000ml of distilled water.
  • Page 8: Cell

    b. CELL The second screen will display CELL and the current cell value. The default cell value is 0.50 and is displayed in the lower right of the screen. The unit allows a variance of +/- 0.15 before displaying an error message. The cell value can’t be adjusted at this screen; calibrating conductivity is the only way to adjust the cell constant.
  • Page 9: Conductivity Measurements

    Decrease the sample Over “over” during > 194°F (90°C) temperature Measurements Temperature is Increase the sample undr < 14°F (-10°C) temperature [Note: If the unit still does not perform normally after the above measures are taken, call ECD Service Department.]...
  • Page 10: Specifications

    Specifications Range Resolution Accuracy Display 0.01uS/cm ±1% of reading + 2uS/cm 0.0 to 499.9uS/cm 1uS/cm ±1% of reading + 5uS/cm 500 to 4999uS/cm Conductivity 0.01mS/cm ±1% of reading + 0.05mS/cm 5.00 to 49.99mS/cm 0.1mS/cm ±2.5% of reading + 0.5mS/cm 50.0 to 200.0mS/cm Salinity 0.0 to 70.0ppt 0.1ppt...
  • Page 11: Warranty

    Seller’s sole liabilities and the buyer’s sole remedies under this agreement shall be limited to a refund in the purchase price, or at ECD’s discretion, to the repair or replacement of any product that proves, upon ECD’s examination, to be defective, when returned to the factory, transportation prepaid by the buyer, within one (1) year of the product’s original shipment date.
  • Page 12 1500 N. Kellogg Dr. Anaheim, CA 92807 Phone: +1 (714) 695-0051 www.ecdanalytical.com...

Table of Contents