Config (Configuration Menu) - ECD TC80 Instruction Manual

Total chlorine analyzer; dechlorination analyzer
Table of Contents

Advertisement

 TEMP allows the displayed temperature to be trimmed to agree with actual process temperature.
3.2.3 CONFIG (C
ONFIGURATION
Four options are available in the Configure Menu, XMTR, SENSOR, LOAD DEFAULT and Dampen.
 XMTR enters the Transmitter Configuration menu.
o LCD access the Display Configuration Menu
o OUTPUT access the Output Configuration Menu
M
)
ENU
SETUP adjust screen lighting characteristics
Temp. Choose °C or °F
CONT adjust Contrast
BACK LIGHT adjust Backlight Timeout, from always
ON to OFF after 10 minutes
GRAPH provides the choice of which Graph style is displayed
on the Home screen.
LINE , Moving average, vertical scale set to 0-100%
of the 4-20 mA output and user defined time scale
GAUGE, Current reading 0-100% of 4-20 mA range
BAR, Current reading 0-100% of 4-20 mA range
LABELS
TAG, Enter up to 2 lines x 16 characters, example,
Name, tag #... Displayed in INFO screen
TAG ON, Turn TAG ON/OFF, adds TAG to Main
Display Sequence, DATA → mV → GRAF → TAG →
DATA
POP UP, Turns ON/OFF, the double tap HOME
Screen pop up memo
SENSOR, Enter up to 2 lines x 16 characters
4-20 mA configure 4-20 mA output (PV or Temp or More)
PV RANGE Enter 4 mA value and 20 mA value
Temp RANGE Enter 4 mA value and 20 mA value
MORE
o CAL Trim 4.00 mA and 20.00 mA output
o FAULT Choose fault condition 3.5 mA, 22
mA, None
RELAY
RLY1,2,3 Choose relay type:
o Alarm, enter the Set point ON, Set Point OFF, Expiration time, Delay ON
and Delay OFF times and the State, energize: changes state from de-
energized to energized on alarm.
o Timed, Enter Period, Duration times and
Hold On/Off
o Fault, No input required, relay condition
changes from energize to de-energize.
24